Comparison of radiation exposure and surgery time between an intraoperative CT with automatic surface registration and a preoperative CT with manual surface registration in navigated spinal surgeries
Martin Mohammed Marzouk, Yama Afghanyar, Mark Mahmoud Marzouk, Sarah Halima Boussouf, Philipp Hartung, Marcus Richter
January 2022, pp 1 - 8 Original Article Read Full Article 10.1007/s00586-021-07082-x
First Online: 07 January 2022
Purpose
This retrospective matched case–control study was conducted to compare two CT based surgery techniques for navigated screw placement in spinal surgery, whether a reduction of radiation exposure and surgery time could be achieved.
Methods
We matched cases treated with an intraoperative CT (iCT), regarding the type and number of implants, with cases treated with a preoperative CT (pCT) of one main surgeon. Outcome measures were radiation exposure due to intraoperative control x-rays, radiation exposure due to CT images, and the duration of surgery.
Results
The required radiation exposure could be significantly reduced in the iCT group. For the intraoperative control X-rays by 69% (median (MED) 88.50/standard deviation (SD) 107.84 and MED 286.00/SD 485.04 for iCT and pCT respectively—in Gycm2; p < 0.001) and for the CT examinations by 25% (MED 317.00/SD 158.62 and MED 424.50/SD 225.04 for iCT and pCT respectively—in mGycm; p < 0.001) with no significant change in surgery time. The correlation between the number of segments fused and the necessary surgery time decreased significantly for the iCT group (Pearson product-moment-correlation: r = 0.569 and r = 0.804 for iCT and pCT respectively; p < 0.05).
Conclusion
The results show that spinal navigation using an intraoperative CT with automatic registration compared to a preoperative CT and intraoperative manual surface registration, allows a significant reduction of radiation exposure, without prolonged surgery time. A significant benefit regarding cut-to-suture-time can be gained with surgeries of a larger scale.
Read Full Article