Konrad Seller, Dieter Wahl, Alexander Wild, Rüdiger Krauspe, Erich Schneider, Berend Linke


July 2007, Volume 16, Issue 7, pp 1047 - 1054 Original Article Read Full Article 10.1007/s00586-007-0307-0

First Online: 02 February 2007

A lot of new implant devices for spine surgery are coming onto the market, in which vertebral screws play a fundamental role. The new screws developed for surgery of spine deformities have to be compared to established systems. A biomechanical in vitro study was designed to assess the bone–screw interface fixation strength of seven different screws used for correction of scoliosis in spine surgery. The objectives of the current study were twofold: (1) to evaluate the initial strength at the bone–screw interface of newly developed vertebral screws (Universal Spine System II) compared to established systems (product comparison) and (2) to evaluate the influence of screw design, screw diameter, screw length and bone mineral density on pullout strength. Fifty-six calf vertebral bodies were instrumented with seven different screws (USS II anterior 8.0 mm, USS II posterior 6.2 mm, KASS 6.25 mm, USS II anterior 6.2 mm, USS II posterior 5.2 mm, USS 6.0 mm, USS 5.0 mm). Bone mineral density (BMD) was determined by quantitative computed tomography (QCT). Failure in axial pullout was tested using a displacement-controlled universal test machine. USS II anterior 8.0 mm showed higher pullout strength than all other screws. The difference constituted a tendency (P = 0.108) when compared to USS II posterior 6.2 mm (+19%) and was significant in comparison to the other screws (+30 to +55%, P 